GUI reference
R&S
®
ZNA
596User Manual 1178.6462.02 ─ 20
Save Verification Data
Same as Export Data to METAS VNA Tools. The exported ZIP file now also contains
the METAS verification configuration, and the verification measurement data obtained
so far.
5.11.5.4 METAS Reconnection dialog
While a METAS calibration is active, you can notify the METAS VNA Tools about addi-
tional reconnections and/or cable movements. The latter is only applicable if cables
were configured during the setup of the METAS calibration.
Access: main toolbar icon or
Channel – [Cal] > "METAS Cal" > "METAS Reconnection"
Both reconnections and cable movements contribute to the overall measurement
uncertainty. In the METAS VNA Tools project created during the METAS calibration),
these events are registered in the so-called measurement journal. In addition, the
results measured immediately after the event report are persisted in the project.
Cal softtool