EasyManua.ls Logo

R&S ZNA67 - Page 897

R&S ZNA67
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
897User Manual 1178.6462.02 ─ 20
Query parameters:
<Data> ALL
If omitted, a single numeric value is returned. If ALL is specified,
the result consists of 6 numeric values. Furthermore, the inter-
pretation of the result values depends on the active trace's stim-
ulus axis (CALCulate<Chn>:TRANsform:TIME:XAXis). See
the table below.
Note that the skew value is positive, if, at the defined position
(CALCulate<Chn>:DTIMe:POSition), the active trace is to
the right of the reference trace.
Usage: Query only
Options: R&S ZNA-K20
Manual operation: See "Extended Info" on page 759
<Data> omitted <Data> = ALL
Stimulus axis:
time
<skew time> <skew time>,<skew distance>, <X value of ref. point on current
trace>, <Y value of ref. point on current trace>, <X value of ref.
point on reference trace>, <Y value of ref. point on reference
trace>
Stimulus axis:
distance
<skew distance> <skew distance>, <skew time>, <X value of ref. point on cur-
rent trace>, <Y value of ref. point on current trace>, <X value of
ref. point on reference trace>, <Y value of ref. point on refer-
ence trace>
CALCulate<Chn>:DTIMe:LIMit:FAIL? <Boolean>
Indicates whether the skew has passed or failed.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Return values:
<Boolean> 0 – skew check has passed
1 – skew check has failed
Usage: Query only
Options: R&S ZNA-K20
CALCulate<Chn>:DTIMe:LIMit:FAIL:BEEP <Boolean>
Defines whether the R&S ZNA should make an audible beep on skew limit violations.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> *RST: OFF
Options: R&S ZNA-K20
SCPI command reference

Table of Contents

Other manuals for R&S ZNA67

Related product manuals