Command reference
R&S
®
ZNA
898User Manual 1178.6462.02 ─ 20
Manual operation: See "Skew Fail Beep" on page 759
CALCulate<Chn>:DTIMe:LIMit:LIMit <arg0>
Defines the limit value for the skew check.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<arg0> Limit value in seconds or meters, depending on the format of the
current trace's stimulus axis (see CALCulate<Chn>:
TRANsform:TIME:XAXis).
Options: R&S ZNA-K20
Manual operation: See "Skew Limit" on page 759
CALCulate<Chn>:DTIMe:LIMit:STATe
<Boolean>
Activates/deactivates the skew limit check.
Use CALCulate<Chn>:DTIMe:LIMit:LIMit to set the applicable limit.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> *RST: OFF
Options: R&S ZNA-K20
Manual operation: See "Skew Check On" on page 759
CALCulate<Chn>:DTIMe:POSition <SkewPos>
Defines the position of the skew measurement.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<SkewPos> Skew position as integer percentage of the step size
Options: R&S ZNA-K20
Manual operation: See "Skew Position" on page 759
CALCulate<Chn>:DTIMe:STATe <Boolean>
Enables/disables the Skew Measurement.