Command reference
R&S
®
ZNA
932User Manual 1178.6462.02 ─ 20
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:USER:COUNt?
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:
USEDefault).
This command returns the number of custom frequencies, defined using CALCulate:
FMODel:DELT<Ph_pt>:FREQuencies:USER:ADD etc.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage: Query only
Options: R&S ZNA-K231
Manual operation: See "Frequencies" on page 700
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:USER:ADD
<UserFrequencies>
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:USER:DELete <Frequency>
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:
USEDefault).
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:USER:ADD|DELete adds/
deletes a frequency to|from the list of custom frequencies.
Suffix:
<Ph_pt>
.
This suffix is ignored
Setting parameters:
<Frequency> Custom frequency
Range: The frequency range of the analyzer
Default unit: Hz
Usage: Setting only
Options: R&S ZNA-K231
Manual operation: See "Frequencies" on page 700
CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:USER:DELete:ALL
The Delta-L algorithm can calculate the loss-per-inch uncertainties either at default or
at custom frequencies (see CALCulate:FMODel:DELT<Ph_pt>:FREQuencies:
USEDefault).
This command clears the list of custom frequencies.
Suffix:
<Ph_pt>
.
This suffix is ignored
Usage: Event
Options: R&S ZNA-K231
SCPI command reference