Wherein,
PhaseA
F
B
F
represents Phase (f-f),
DelayA
F
B
F
represents Delay(f-f), and
Period
sourceA
represents the period of Source A.
7. Phase(r-f): indicates the phase deviation between the threshold middle values of
the rising edge of Source A and the falling edge of Source B. The phase formula is
as follows:
°360×=
sourceA
FR
FR
Period
BDelayA
BPhaseA
Wherein,
PhaseA
R
B
R
represents Phase (r-f),
DelayA
R
B
R
represents Delay(r-f), and
Period
sourceA
represents the period of Source A.
8. Phase(f-r): indicates the phase deviation between the threshold middle values of
the falling edge of Source A and the rising edge of Source B. The phase formula is
as follows:
°360×=
sourceA
RF
RF
Period
BDelayA
BPhaseA
Wherein,
PhaseA
F
B
R
represents Phase (f-r),
DelayA
F
B
R
represents Delay(f-r), and
Period
sourceA
represents the period of Source A.
TIP
• Source A and Sour
ce B can be any channel among CH1~CH4 and Math1~Math4.
• The default threshold middle value is 50%.
10.2.1.4 Voltage Parameters
Vpp
Vtop
Vbase
Overshoot
Vmax
Vmin
Vupper
Vmid
Vlower
Preshoot
Figure 10.3 Voltage Parameters
Measurements
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
DHO1000 User Guide
143
www.calcert.com sales@calcert.com1.888.610.7664
0
5
10
15
20
25
30

Table of Contents

Other manuals for Rigol DHO1000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DHO1000 Series and is the answer not in the manual?

Rigol DHO1000 Series Specifications

General IconGeneral
BrandRigol
ModelDHO1000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals