SWEEP Key - Receiver R&S ESCI
1166.6004.12 4.48 E-2
Scan Stop - stop frequency of display range
Range is f
min
to f
max
.
his value can also be set in the FREQ menu (STOP softkey).
IEC/IEEE-bus command FREQ:STOP 1250 MHz
Step Mode - selection of frequency switching mode
Linear or logarithmic frequency switching can be selected. The selected
setting is valid for all scan ranges.
STEP MODE
LIN
LOG
AUTO
LIN linear frequency switching.
AUTO linear frequency switching.
The step width is selected automatically depending on the set
resolution bandwidth so that all signals occurring in the scan
range are reliably detected without any significant
measurement error (about one third of resolution bandwidth)
LOG logarithmic frequency switching.
The frequency is incremented in % of the current frequency..
IEC/IEEE-bus command SWE:SPAC LIN
Start - Entry of start frequency
The start frequency of a subrange must be equal to or greater than the stop
frequency of the previous subrange.
On entering the start frequency, the preceding scan range is – if necessary –
adapted automatically to avoid overlapping of scan ranges.
IEC/IEEE-bus command SCAN1:STAR 150 kHz
Stop - Entry of stop frequency
The stop frequency of a subrange must be equal to or greater than the start
frequency of the subrange.
On entering the stop frequency, the preceding scan range is – if necessary –
adapted automatically to avoid overlapping of scan ranges.
IEC/IEEE-bus command SCAN1:STOP 30 MHz