TRACE - Receiver R&S ESCI
1166.6004.12 4.64 E-2
BLANK
The BLANK softkey activates the blanking of the trace on the screen.
IEC/IEEE-bus command DISP:WIND:TRAC OFF
SCAN
COUNT
The SCAN COUNT softkey activates the entry of the number of scan used in
the SINGLE SCAN mode.
The allowed range of values is 0 to 30000. The default setting is 1.
IEC/IEEE-bus command SWE:COUN 10
DETECTOR
The DETECTOR softkey opens a submenu to select the detector.
The detector type may be indecently selected for each trace.
For description of detector types see Section "Detector Selection" in Analyzer
Mode.
MAX PEAK
The MAX PEAK softkey activates the max peak detector.
IEC\IEEE bus command DET POS
MIN PEAK
The MIN PEAK softkey activates the min peak detector.
IEC\IEEE bus command DET NEG
QUASIPEAK
The QUASIPEAK softkey selects the quasi-peak detector.
The IF bandwidth is adapted as a function of the frequency
range. The coupling of the IF bandwidth to the frequency
range can be cancelled using softkey QP RBW
UNCOUPLED.
IEC/IEEE-bus command DET QPE
AVERAGE
The AVERAGE softkey activates the average detector.
IEC\IEEE bus command DET AVER
RMS
The RMS softkey activates the rms detector.
IEC\IEEE bus command :DET RMS