R&S ESCI Analyzer - Sweep
1166.6004.12 4.91 E-1
CONTINUE
SGL SWEEP
The CONTINUE SGL SWEEP softkey repeats the number of sweeps set
under SWEEP COUNT, however without first deleting the trace.
This is particularly of interest when using the functions TRACE AVERAGE
and MAXHOLD, if previously recorded measurement results are to be taken
into consideration for averaging / maximum search.
If SGL SWEEP DISP OFF is active, the screen is switched off also during
repeated sweeps.
IEC/IEEE-bus command: INIT:CONM
SWEEPTIME
MANUAL
The SWEEPTIME MANUAL softkey activates the window for entering the
sweep time manually (see also BW menu).
IEC/IEEE-bus command: SWE:TIME 10s
SWEEPTIME
AUTO
The SWEEPTIME AUTO softkey activates the automatic selection of the
sweep time as a function of the bandwidth of the resolution and video filters
(see also BW menu).
IEC/IEEE-bus command: SWE:TIME:AUTO ON
SWEEP
COUNT
The SWEEP COUNT softkey activates the window for the entry of the
number of sweeps to be performed by R&S ESCI after a single sweep has
been started. If Trace Average, Max Hold or Min Hold is activated, this also
determines the number of averaging or maximum search procedures.
Example:
[TRACE1: MAX HOLD]
[SWEEP: SWEEP COUNT: {10} ENTER]
[SINGLE SWEEP]
R&S ESCI performs the Max Hold function over 10 sweeps.
The permissible range for the sweep count is 0 to 32767. For sweep count =
0 or 1, one sweep is performed. For trace averaging in the continuous-sweep
mode, R&S ESCI performs running averaging over 10 sweeps if sweep count
= 0; if sweep count = 1, no averaging is performed.
The sweep count is valid for all the traces in a diagram.
Note: The number of sweeps set in the TRACE menu is the same as
that in the SWEEP menu.
If SINGLE SWEEP is selected, the measurement stops after the
selected number of sweeps has been performed.
IEC/IEEE-bus command: SWE:COUN 64