R&S ESCI Configuration/Setup
1166.6004.12 4.215 E-3
FREQUENCY, TDF/dB – Input of reference values
The selection bar marks the first reference value. The desired reference
values have to be entered in ascending frequency sequence.
After entering the frequency the selection bar automatically jumps to the
corresponding level value.
After entering the first reference value the table can be edited. The two
INSERT LINE and DELETE LINE softkeys are displayed. Individual values
are changed at a later time by marking the field and by entering the new
value.
IEC/IEEE-bus command CORR:TRAN:DATA <freq>,<level>,...
INSERT
LINE
The INSERT LINE softkey inserts a free line above the
marked reference value. When entering a new reference
value into this line watch out for the ascending frequency
sequence.
DELETE
LINE
The DELETE LINE softkey deletes the marked reference
value (the whole line). The following reference values move
up.
IEC/IEEE-bus command --
SAVE TRD
FACTOR
The SAVE TRD FACTOR softkey saves the changed table
in a file on the internal hard disk.
If a transducer factor with the same name already exists, a
corresponding query is performed beforehand.
If the factor to be newly saved is currently switched on, the
new values will immediately be valid. If a transducer set is
switched on comprising the factor, the values will only be
used when the set is switched on next time.
IEC/IEEE-bus command - (executed automatically
after definition of reference values)