Configuration/Setup R&S ESCI
1166.6004.12 4.214 E-3
Unit – Selection of unit
The unit of the transducer factor is selected from a selection box activated by
the ENTER key.
FACTOR U
N
IT
dB
dBm
dBµV
dBµV/m
dBµA
dBµA/m
dBp
W
dBpT
The default setting is dB.
IEC/IEEE-bus command CORR:TRAN:UNIT <string>
Interpolation - Selection of interpolation
Linear or logarithmic interpolation can be carried out between the frequency
reference points of the table. Selection is via the ENTER key which is toggled
between LIN and LOG (toggle function).
IEC/IEEE-bus command CORR:TRAN:SCAL LIN|LOG
The following diagrams indicate the effect of the interpolation on the
calculated characteristic:
Fig. 4-22 Linear frequency axis and linear interpolation
Fig. 4-23 Logarithmic frequency axis and logarithmic interpolation
Comment - Entry of a comment
The comment can be freely selected. It can have a maximum number of 50
characters.
IEC/IEEE-bus command CORR:TRAN:COMM <string>