General Instrument Settings and Instrument Setup
R&S
®
SGS100A
128User Manual 1173.9105.02 ─ 12
8 General Instrument Settings and Instrument
Setup
This section describes the settings which do not directly affect signal generation.
Access:
► Select "SGMA-GUI > Instrument Name > Setup" and select the required dialog.
8.1 Internal Adjustments
The R&S SGS is accurate due to the integrated procedures for adjustments.
All internal adjustments for which no external measuring equipment is needed can be
started in the "Internal Adjustments" dialog. The adjustments with external measuring
equipment are described in the service manual.
Adjustment is recommended if the temperature range in which the instrument is oper-
ated changes, before all applications which require maximum level and frequency
accuracy, or after enabling/disabling of the "Eco Mode" (see Chapter 8.12, "Eco Mode",
on page 143).
During adjustment, a bar indicates the status of progress. If an error occurs, adjust-
ment is terminated and an error message is output in the info line.
Risk of invalid adjustment
Wait until the instrument has reached its operating temperature before you start the
adjustment procedure. The warm-up time is 30 minutes.
Internal Adjustments