Remote Control Commands
R&S
®
SGS100A
205User Manual 1173.9105.02 ─ 12
Activating the original state of delivery (factory reset, :SYSTem:FPReset
on page 205). Only functions that are protected by a password remain unchanged
as well as the passwords themselves.
:SOURce<hw>:PRESet
:SYSTem:PRESet
Triggers an instrument reset. It has the same effect as:
The *RST command
The "SGMA-GUI > Instrument Name > Preset" function
For an overview of the settings affected by the preset function, see Chapter 7.9, "Pre-
set", on page 123.
Example:
SYST:PRES
All instrument settings (also the settings that are not currently
active) are reset to their default values.
Usage: Setting only
:SYSTem:FPReset
Triggers an instrument reset to the original state of delivery.
Example:
SYST:FPR
all instrument settings (also those that are not currently active)
are reset to the factory values.
Usage: Event
Manual operation: See "Execute Factory Preset" on page 143
11.5 CALibration Subsystem
:CALibration:ALL[:MEASure]?......................................................................................... 205
:CALibration:FREQuency[:MEASure]?..............................................................................206
:CALibration:IQModulator:FULL?..................................................................................... 206
:CALibration:IQModulator:LOCal?....................................................................................206
:CALibration:LEVel[:MEASure]?.......................................................................................206
:CALibration:LEVel:TEMPerature?................................................................................... 207
:CALibration:FREQuency:TEMPerature?..........................................................................207
:CALibration:IQModulator:TEMPerature?..........................................................................207
:CALibration:OEXTension................................................................................................207
:CALibration:ROSCillator:DATA:MODE.............................................................................207
:CALibration:ROSCillator[:DATA]......................................................................................208
:CALibration:ALL[:MEASure]?
Starts all internal adjustments for which no external measuring equipment is needed.
CALibration Subsystem