Signal Generator Settings
R&S
®
SGS100A
90User Manual 1173.9105.02 ─ 12
Limit
Sets the level limit.
The value specifies the upper limit of the level at the [RF Out] connector. A message
appears if an attempt is made to set a level above this limit and the level at the RF out-
put is confined to the upper limit. However, the level indication is not influenced.
The value is not affected by an instrument preset function. This parameter is influenced
only by the Factory Preset and its factory value is equal to the upper limit.
Remote command:
[:SOURce]:POWer:LIMit[:AMPLitude] on page 252
Readjust
Recalculates the instrument internal settings optimized for the current level. Not
required for automatic modes.
Remote command:
[:SOURce]:POWer:ALC:SONCe on page 249
7.5.2 Attenuator
To access the attenuator settings, select "SGMA-GUI > Instrument Name > Level
>Attenuator".
This dialog comprises the settings for the power-on behavior of the instrument.
Mode
Sets the attenuator mode at the RF output.
"Auto"
Standard mode.
The electronically switching attenuator switches with a ~ 6 dB step
width at optimized switching points. The entire level range is availa-
ble.
The level setting is performed by continuous electronic level control
combined with switching the step attenuator.
Level and Power-On Settings