Remote Control Commands
R&S
®
SGS100A
272User Manual 1173.9105.02 ─ 12
Parameters:
<Parity> NONE | ODD | EVEN
*RST: n.a. (factory preset: NONE)
Example:
SYST:COMM:SER:PAR NONE
Selects parity NONE.
:SYSTem:COMMunicate:SERial:RESource?
Queries the VISA resource string for the serial remote control interface. This string is
used for remote control of the instrument.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:SERial:RESource?
queries the VISA resource string.
Response: "ASRL1::INSTR"
Usage: Query only
:SYSTem:COMMunicate:SERial:SBITs <SBits>
Sets the number of stop bits for the serial remote control interface.
Parameters:
<SBits> 1 | 2
*RST: n.a. (factory preset: 1)
Example:
SYST:COMM:SER:SBIT 2
Selects 2 stop bits.
:SYSTem:COMMunicate:HISLip:RESource?
Queries the VISA resource string, used for remote control of the instrument with
HiSLIP protocol.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:HISLip:RESource?
Response: "TCPIP::192.1.2.3::hislip0::INSTR"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 142
:SYSTem:COMMunicate:PCIexpress:RESource?
Queries the visa resource string for remote control via the PCIe interface.
Return values:
<Resource> string
SYSTem Subsystem