General Instrument Settings and Instrument Setup
R&S
®
SGS100A
130User Manual 1173.9105.02 ─ 12
I/Q Modulator
Starts the adjustment for the I/Q modulator for the entire frequency range. The I/Q
modulator is adjusted regarding carrier leakage, I/Q imbalance and quadrature.
Remote command:
:CALibration:IQModulator:FULL? on page 206
Delta Temperature
Displays the difference between the current temperature and the temperature by the
last performed adjustment.
Note: Adjustment is recommended if the temperature range in which the instrument is
operated changes, before all applications which require maximum level and frequency
accuracy, or after enabling/disabling of the "Eco Mode" (see Chapter 8.12, "Eco Mode",
on page 143).
Remote command:
:CALibration:LEVel:TEMPerature? on page 207
:CALibration:FREQuency:TEMPerature? on page 207
:CALibration:IQModulator:TEMPerature? on page 207
8.2 Hardware Configuration
Querying information about the installed assemblies
Select "SGMA-GUI > Instrument Name > Setup > Hardware Config".
In the "Hardware Config" dialog, the installed assemblies together with their var-
iants and revision states can be displayed for servicing purposes.
Hardware Configuration