Remote Control Commands
R&S
®
SGS100A
229User Manual 1173.9105.02 ─ 12
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 225.
Manual operation: See "Default Aperture Time" on page 111
:SENSe<ch>[:POWer]:APERture:TIMe <ApTime>
Defines the aperture time (size of the acquisition interval) for the corresponding sensor.
Parameters:
<ApTime> float
Range: depends on connected power sensor
Increment: 1E-9
*RST: depends on connected power sensor
Example: See Example "Performing a power measurement with a fixed fil-
ter" on page 225.
Manual operation: See "Aperture Time" on page 112
:SENSe<ch>[:POWer]:CORRection:SPDevice:SELect <Select>
Several S-parameter tables can be stored in a sensor. The command selects a loaded
data set for S-parameter correction for the corresponding sensor.
Parameters:
<Select> float
*RST: 0
Manual operation: See " S-Parameter " on page 112
:SENSe<ch>[:POWer]:CORRection:SPDevice:STATe <State>
Activates the use of the S-parameter correction data.
Note: If you use power sensors with attenuator, the instrument automatically activates
the use of S-parameter data.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
SENSe1:POWer:CORRection:SPDevice:STATe 1
Activates the use of the S-parameters correction data.
Manual operation: See " S-Parameter " on page 112
:SENSe<ch>[:POWer]:CORRection:SPDevice:LIST?
Queries the list of the S-parameter data sets that have been loaded to the power sen-
sor.
SENSe, READ, INITiate and SLISt Subsystems