Remote Control Commands
R&S
®
SGS100A
204User Manual 1173.9105.02 ─ 12
0
test query to prove whether the instrument is locked
Controller ID
request lock from the controller with the specified Controller ID
Return values:
<Value> Number
0
request refused; the instrument is already locked to other <Lock
Request Id>, i.e. to another controller
1
request granted
Example:
:LOCK? 12345
Response: 1
:UNL 12345
Usage: Query only
Manual operation: See "Exclusive Access" on page 58
:UNLock <UnlockId>
Unlocks an instrument locked to a controller with Controller ID = <Unlock Id>.
Setting parameters:
<UnlockId> Number
Unlock ID which uniquely identifies the controller to the instru-
ment. The value must match the Controller ID <Lock Request
Id> set with the command :LOCK?.
0
Clear lock regardless of locking state
Usage: Setting only
Manual operation: See "Exclusive Access" on page 58
11.4 Preset Commands
The preset commands are not bundled into one subsystem. Therefore, they are listed
separately in this section. In addition, a specific preset command is provided for each
digital standard and for the fader. These specific commands are described in the asso-
ciated subsystems.
The following presetting actions are available:
●
Activating the default state of all internal instrument functions (*RST on page 202).
Functions that concern the integration of the instrument into a measurement setup
are not changed, e.g. TCP/IP address or reference oscillator source settings.
Preset Commands