Performing Configuration Tasks
R&S
®
SGS100A
155User Manual 1173.9105.02 ─ 12
3. Avoid unnecessary cable lengths and branching points.
Refer to Figure 9-3 for an example of how to connect two instruments for achieving
phase coherence. The configuration can be extended by further instruments.
Figure 9-3: Example of a setup: LO coupling
To configure the first instrument in the chain
1. In the "Frequency / Phase > Local Oscillator (LO) Coupling" dialog, select "LO
Source > Internal".
2. In the "Frequency / Phase > Local Oscillator (LO) Coupling" dialog, select "REF/LO
Output > LO".
3. If an external reference frequency is used, select "SGMA-GUI > Instrument Name
> Reference Oscillator > Source > External".
To configure next instruments
1. In the "Frequency / Phase > Local Oscillator (LO) Coupling" dialog, select "LO
Source > External".
2. In the "Frequency / Phase > Local Oscillator (LO) Coupling" dialog, select "REF/LO
Output > LO".
9.8 How to Restore the LAN Connection to an Instrument
► If the LAN connection to an instrument configured to use a static IP address is lost,
press the [LAN LED] on the instrument's front panel for more than 3 seconds to
reset the LAN settings and to set the "Address Mode" on page 140 to DHCP.
Provided the network supports automatic assignment of IP address, new IP
address is automatically assigned to the instrument.
How to Restore the LAN Connection to an Instrument