Remote Control Commands
R&S
®
SGS100A
235User Manual 1173.9105.02 ─ 12
:SENSe<ch>[:POWer]:TYPE?
Queries the sensor type. The type is automatically detected.
Return values:
<Type> string
Example:
SENS1:TYPE?
Queries the type of sensor.
Response: NRP-Z21
The R&S NRP-Z21 sensor is used.
Usage: Query only
Manual operation: See " Sensor type and serial number " on page 109
:SENSe<ch>[:POWer]:ZERO
Performs zeroing of the sensor.
Zeroing is required after warm-up, i.e. after connecting the sensor.
Note: Switch off or disconnect the RF power source from the sensor before zeroing.
We recommend that you zero in regular intervals (at least once a day), if:
●
The temperature has varied more than about 5 °C.
●
The sensor has been replaced.
●
You want to measure very low power.
Example:
SENS1:ZERO
Executes zeroing.
Usage: Event
Manual operation: See " Zero " on page 109
11.13 SOURce Subsystem
[:SOURce]:OPMode....................................................................................................... 235
[:SOURce]:FREQuency[:CW|FIXed].................................................................................236
[:SOURce]:FREQuency:OFFSet......................................................................................236
[:SOURce]:LOSCillator:SOURce......................................................................................236
[:SOURce]:PATH:COUNt?...............................................................................................237
[:SOURce]:OPMode <OpMode>
Sets the operation mode.
Parameters:
<OpMode> NORMal | BBBYpass
NORMal
normal operation
SOURce Subsystem