11.11 OUTPut Subsystem
:OUTPut:AFIXed:RANGe:LOWer?...................................................................................222
:OUTPut:AFIXed:RANGe:UPPer?....................................................................................222
:OUTPut:AMODe........................................................................................................... 222
:OUTPut[:STATe]:PON....................................................................................................223
:OUTPut[:STATe]............................................................................................................223
:OUTPut:AFIXed:RANGe:LOWer?
Queries the minimum level which can be set without the attenuator being adjusted
(Attenuator FIXed).
Return values:
<Lower> float
Default unit: dBm
Example: See Chapter 11.1.2, "Generating an I/Q Modulated Signal",
on page 193.
Usage: Query only
Manual operation: See "Level Range" on page 91
:OUTPut:AFIXed:RANGe:UPPer?
Queries the maximum level which can be set without the attenuator being adjusted
(Attenuator FIXed).
Return values:
<Upper> float
Default unit: dBm
Example: See Chapter 11.1.2, "Generating an I/Q Modulated Signal",
on page 193.
Usage: Query only
Manual operation: See "Level Range" on page 91
:OUTPut:AMODe <AMode>
Switches the mode of the attenuator at the RF output.
Parameters:
<AMode> AUTO | FIXed | APASsive
AUTO
The attenuator is switched automatically. The level settings are
made in the full range.
OUTPut Subsystem