EasyManua.ls Logo

Rohde & Schwarz SGS100A - 11.17 Source:power Subsystem; Source:power Subsystem; [:Source]:Power:alc:dsensitivity; [:Source]:Power:alc:sonce

Rohde & Schwarz SGS100A
324 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
SGS100A
249User Manual 1173.9105.02 ─ 12
11.17 SOURce:POWer Subsystem
[:SOURce]:POWer:ALC:DSENsitivity............................................................................... 249
[:SOURce]:POWer:ALC[:STATe]...................................................................................... 249
[:SOURce]:POWer:ALC:SONCe......................................................................................249
[:SOURce]:POWer:ATTenuation:RFOFf:MODE................................................................. 250
[:SOURce]:POWer:ATTenuation:SOVer[:OFFSet]..............................................................250
[:SOURce]:POWer:LMODe............................................................................................. 250
[:SOURce]:POWer:POWer..............................................................................................250
[:SOURce]:POWer:SCHaracteristic..................................................................................251
[:SOURce]:POWer[:LEVel][:IMMediate][:AMPLitude]..........................................................251
[:SOURce]:POWer[:LEVel][:IMMediate]:OFFSet................................................................ 252
[:SOURce]:POWer:LIMit[:AMPLitude]...............................................................................252
[:SOURce]:POWer:PEP?................................................................................................ 252
[:SOURce]:POWer:RANGe:LOWer?................................................................................ 253
[:SOURce]:POWer:RANGe:UPPer?.................................................................................253
[:SOURce]:POWer:ALC:DSENsitivity <Sensitivity>
Sets the power detector sensitivity. Used for compatibility reasons only.
Parameters:
<Sensitivity> OFF | LOW | MED | HIGH
*RST: OFF
Manual operation: See "Detector Sensitivity" on page 93
[:SOURce]:POWer:ALC[:STATe] <State>
Activates/deactivates automatic level control.
Parameters:
<State> 1 | OFFTable | OFF | ONTable | AUTO | ON
*RST: ONTable
Example: See Chapter 11.1.2, "Generating an I/Q Modulated Signal",
on page 193.
Manual operation: See "State" on page 93
[:SOURce]:POWer:ALC:SONCe
Briefly activates level control for correction purposes.
Usage:
Event
Manual operation: See "Readjust" on page 90
SOURce:POWer Subsystem