Remote Control Commands
R&S
®
SGS100A
221User Manual 1173.9105.02 ─ 12
:MMEMory:STORe:STATe <savrcl_state_nr>, <file_name>
Stores the current instrument setting in the specified file.
The instrument setting must first be stored in an internal memory with the same num-
ber using the common command *SAV.
Setting parameters:
<savrcl_state_nr> Corresponds to the specific <number> defined with the *SAV
command, e.g. *SAV 4.
<file_name> String parameter to specify the file name with extension
*.savrcltxt.
Example: See "Storing and Loading Current Settings" on page 214.
Usage: Event
11.10 Fast Speed Commands
This section describes special commands that allow a fast frequency and level setting.
:FFASt.......................................................................................................................... 221
:PFASt.......................................................................................................................... 221
:FFASt <Freq>
Special command to set the RF output frequency with minimum latency. No unit (e.g.
Hz) allowed.
Bypasses the status system so command *OPC? cannot be appended.
Parameters:
<Freq> float
Example:
FFASt 12750000000
:PFASt <Pow>
Special command to set the RF output level with minimum latency at the RF output
connector. This value does not consider a specified offset. No unit (e.g. dBm) allowed.
Bypasses the status system so command *OPC? cannot be appended.
Parameters:
<Pow> float
Example:
:PFASt -20
Fast Speed Commands