Remote Control Commands
R&S
®
SGS100A
207User Manual 1173.9105.02 ─ 12
Return values:
<Level> 0 | 1 | OFF | ON
Example: See Chapter 11.1.1, "Performing General Task for Instrument
Setup", on page 191.
Usage: Query only
Manual operation: See "Level" on page 129
:CALibration:LEVel:TEMPerature?
:CALibration:FREQuency:TEMPerature?
:CALibration:IQModulator:TEMPerature?
Queries the delta temperature since the last performed adjustment.
Return values:
<Temperature> string
Usage: Query only
Manual operation: See "Delta Temperature" on page 130
:CALibration:OEXTension <OExtension>
This SCPI command is used when the instrument is in extension mode.
Enables you to run an internal adjustment only on the extension instrument.
Parameters:
<OExtension> 0 | 1 | OFF | ON
*RST: 0
Example:
:CAL:OEXT ON
activates the mode for calibrating only the extension instrument
:CAL:LEV:MEAS?
calibrates only the extension instrument
:CAL:OEXT OFF
deactivates the mode for calibrating only the extension instru-
ment
:CALibration:ROSCillator:DATA:MODE <Mode>
Defines whether the factory provided or a custom defined calibration value is used to
adjust the reference oscillator.
Parameters:
<Mode> FACTory | CUSTomer
*RST: FACTory
Example:
:CALibration1:ROSCillator:DATA:MODE CUSTomer
:CALibration1:ROSCillator:DATA 35600
Manual operation: See "Active Adjustment Data" on page 146
CALibration Subsystem