Remote Control Commands
R&S
®
SGS100A
269User Manual 1173.9105.02 ─ 12
Return values:
<Serror> string
Example:
SYST:SERR
queries all errors existing in the error queue.
Example:
Response: -221, 'Settings conflict', 153,
'Input voltage out of range'
The two returned errors have occurred since the error queue
was last queried.
Usage: Query only
:SYSTem:VERSion?
Queries the SCPI version the instrument's command set complies with.
Return values:
<Version> string
Example:
SYST:VERS
queries the SCPI version.
Response: "1996"
The instrument complies with the SCPI version from 1996.
Usage: Query only
:SYSTem:COMMunicate:NETWork:IPADdress <IpAddress>
Sets the IP address.
Parameters:
<IpAddress> string
Range: 0.0.0.0. to ff.ff.ff.ff
Example:
SYSTem:COMMunicate:NETWork:IPADdress "7.8.9.10"
sets the IP address of the instrument.
Manual operation: See "IP Address" on page 141
:SYSTem:COMMunicate:NETWork:IPADdress:MODE <Mode>
Selects manual or automatic setting of the IP address.
Parameters:
<Mode> AUTO | STATic
*RST: n.a. (factory preset: AUTO)
Example:
SYSTem:COMMunicate:NETWork:IPADdress:MODE AUTO
The IP address is assigned automatically (DHCP)
Manual operation: See "Address Mode" on page 140
SYSTem Subsystem