General Instrument Settings and Instrument Setup
R&S
®
SGS100A
141User Manual 1173.9105.02 ─ 12
Note: Lost LAN connection to an instrument.
If the connection to an instrument configured to use static IP addresses is lost, press
the [LAN LED] on the instrument front panel.
This resets the "Address Mode" to its default value ("Auto (DHCP)").
"Auto (DHCP)"
The IP address is assigned automatically.
The network used must support automatic assignment of IP address
via DHCP or APIPA (Zeroconf) to use this function.
"Static"
The IP address is assigned manually.
Remote command:
:SYSTem:COMMunicate:NETWork:IPADdress:MODE on page 269
IP Address
Displays the IP address. To enter the IP address manually, select "Address Mode >
Static".
If there is manual input of the IP address, it is recommended that a connection of the
instrument to the network is coordinated with the network administrator. Connection
errors may affect the entire network.
Remote command:
:SYSTem:COMMunicate:NETWork:IPADdress on page 269
Subnet Mask
Displays the subnet mask. To enter the subnet mask manually, select "Address Mode >
Static".
This number is used together with the IP address to identify the network segment the
instrument is in.
It is recommended that a connection of the instrument to the network is coordinated
with the network administrator. Connection errors may affect the entire network.
Remote command:
:SYSTem:COMMunicate:NETWork[:IPADdress]:SUBNet:MASK on page 271
Default Gateway
Displays the IP address of the default gateway. To enter the default gateway manually,
select "Address Mode > Static".
This address identifies the router on the same network as the instrument that is used to
forward traffic to destinations beyond the local network.
It is recommended that a connection of the instrument to the network is coordinated
with the network administrator. Connection errors may affect the entire network.
Remote command:
:SYSTem:COMMunicate:NETWork[:IPADdress]:GATeway on page 270