Network and Remote Control Operation
R&S
®
SGS100A
177User Manual 1173.9105.02 ─ 12
Permitted PCIe cables
PCIe extension cables must fulfill the following requirements:
Single lane connectors
Max. cable length of 5 m.
For example: OSS-PCIe-CBL-x1 cable from One Stop Systems or 74576-000x cable
from Molex.
Connecting an external PC that does not support hot-plugging
1. Switch off the external PC and the instrument.
See also Chapter 2.1.5, "Switching the Instrument On and Off", on page 19.
2. Connect the instrument and the controller with the suitable cable.
3. Switch on the instrument.
4. Wait until the instrument has completed the booting (the "POWER" LED on the
instrument's front panel is constantly on).
5. Switch on the external PC.
10.3.5 Enabling Fast Settings
To enable the special PCI express feature fast settings, enable the function
rssgs_UseFastSettings (ViSession instrumentHandle, ViBoolean
fastEnabled, ViBoolean asynchronousEnabled) included in the instru-
ment driver.
Settings for some parameters like level and frequency accelerate.
To disable the fast settings, call the function rssgs_UseFastSettings with argu-
ment fastEnabled=false.
10.4 Advanced Remote Control Using Fast Socket
Fast settings
The socket interface can be utilized not only to transfer text messages e.g. SCPI com-
mands but also to carry register based remote control messages. The latter mode is
called fast settings mode. The fast socket communication is based on the Ethernet pro-
tocol which does not support routing. Therefore a controller PC can only control devi-
ces within its own network segment using the fast socket method.
Advanced Remote Control Using Fast Socket