Remote Control Commands
R&S
®
SGS100A
273User Manual 1173.9105.02 ─ 12
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 142
:SYSTem:COMMunicate:SOCKet:PORT <ScpiEthPort>
Sets the port number for remote control via socket communication.
Parameters:
<ScpiEthPort> integer
Range: 1000 to 65535
*RST: n.a. (factory preset: 5025)
Example:
SYSTem:COMMunicate:SOCKet:PORT 5030
// specifies the socket port number.
Manual operation: See "SCPI Port" on page 61
:SYSTem:COMMunicate:SOCKet:RESource?
Queries the VISA resource string for remote control via LAN interface, using TCP/IP
socket protocol.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:SOCKet:RESource?
Response: "TCPIP::10.113.1.150::5025::SOCKET"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 142
:SYSTem:COMMunicate:USB:RESource?
Queries the VISA resource string for remote control via the USB interface.
Return values:
<Resource> string
Example:
SYSTem:COMMunicate:USB:RESource?
queries the VISA resource string for remote control via the USB
interface.
Response: "USB::72::000000::INSTR"
Usage: Query only
Manual operation: See "Visa Resource Strings" on page 142
:SYSTem:REBoot
Restarts the firmware and the operating system.
Usage:
Event