Remote Control Commands
R&S
®
SGS100A
228User Manual 1173.9105.02 ─ 12
Return values:
<Power> float or float,float
The sensor returns the result in the unit set with command :
SENSe<ch>:UNIT[:POWer]
Certain power sensors, such as the R&S NRP-Z81, return two
values, first the value of the average level and - separated by a
comma - the peak value.
Example:
SENS1:UNIT DBM
// Selects unit dBm for presentation of measurement result
READ1?
// Queries the measurement result of the sensor
// -45.6246576745440230
// -45.6 dBm were measured at the given frequency
Example: R&S NRP-Z81
READ1?
-55.62403263352178,-22.419472478812476
-55.6 dBm is the measured average level, -22. 4 dBm is the measured peak level at the given frequency
Usage: Query only
Manual operation: See " Level (Peak) / Level (Average) " on page 109
:SENSe<ch>:UNIT[:POWer] <Power>
Selects the unit (Watt, dBm or dBμV) of measurement result display, queried with :
READ<ch>[:POWer]?.
Parameters:
<Power> DBM | DBUV | WATT
*RST: DBM
Example:
SENS2:UNIT DBM
Selects dBm as unit for the measured value returned by com-
mand READ.
READ2?
Response: 7.34
7.34 dBm are measured by sensor 2.
Manual operation: See " Level (Peak) / Level (Average) " on page 109
:SENSe<ch>[:POWer]:APERture:DEFault:STATe <UseDefAp>
Deactivates the default aperture time of the respective sensor.
To specify a user-defined value, use the command :SENSe<ch>[:POWer]:
APERture:TIMe on page 229.
Parameters:
<UseDefAp> 0 | 1 | OFF | ON
*RST: 1
SENSe, READ, INITiate and SLISt Subsystems