Remote Control Commands
R&S
®
SGS100A
259User Manual 1173.9105.02 ─ 12
EGATe
The pulse modulation is gated by an external gate signal. The
trigger signal is supplied via the trigger connector.
*RST: AUTO
Example:
PULM:TRIG:MODE EXT
selects triggering by an external trigger event.
Manual operation: See "Trigger Mode" on page 117
[:SOURce<hw>]:PULM:WIDTh
<Width>
Sets the width of the generated pulse. The width determines the pulse length. The
pulse width must be at least 20ns less than the set pulse period.
Parameters:
<Width> float
Range: 20 ns to 100 s
Increment: 10 ns
*RST: 2 us
Example:
PULM:WIDT 33 us
sets a width of 33 us for the pulse.
Manual operation: See "Pulse Width" on page 115
11.19 SOURce:ROSCillator Subsystem
[:SOURce]:ROSCillator:EXTernal:FREQuency.................................................................. 259
[:SOURce]:ROSCillator:OUTPut:FREQuency....................................................................260
[:SOURce]:ROSCillator:SOURce..................................................................................... 260
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue.............................................................. 260
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe].............................................................260
[:SOURce]:ROSCillator:EXTernal:SBANdwidth..................................................................261
[:SOURce]:ROSCillator:EXTernal:FREQuency <ExtFreq>
Selects the frequency of the external reference.
Parameters:
<ExtFreq> 10MHZ | 100MHZ | 1000MHZ | 13MHZ
13MHZ requires RF board with part number 1419.5308.02.
To find out the RF board installed in the instrument:
Select "SGMA-GUI > instrument name > Setup > Hardware
Config" > "RF Assembly"
Observe the part number of the assembly "RfBoard".
Example: See Chapter 11.1.4, "Advanced Task for Optimizing Perform-
ance", on page 196.
Manual operation: See "Ext. Ref. Input Frequency" on page 85
SOURce:ROSCillator Subsystem