Signal Generator Settings
R&S
®
SGS100A
99User Manual 1173.9105.02 ─ 12
The correction list entries are only computed when the "Fill" button is pressed.
From
Sets the start value of the index range.
Remote command:
n.a.
Range
Sets the range for filling the table.
Remote command:
n.a.
Select column to fill
Selects either the frequency or the level column to be filled with the value defined
below.
Remote command:
n.a.
Start value
Sets the start value for the frequency or the level entries.
Remote command:
n.a.
End value
Displays the end value for the frequency or the level entries.
Remote command:
n.a.
Increment value
Sets the increment for the frequency or the level entries.
Remote command:
n.a.
Fill
Fills the selected column in the set range with values, starting with the start value and
using the set increment.
Remote command:
n.a.
7.5.7 Fill with Sensor
► To access this dialog, select "SGMA-GUI > Instrument Name > Level > UCOR >
Edit User Cor. Data> Fill With Sensor".
Level and Power-On Settings