Remote Control Commands
R&S
®
SGS100A
258User Manual 1173.9105.02 ─ 12
[:SOURce<hw>]:PULM:TRIGger:EXTernal:GATE:POLarity <Polarity>
Selects the polarity of the Gate signal.
Parameters:
<Polarity> NORMal | INVerted
*RST: NORMal
Example:
PULM:TRIG:EXT:GATE:POL NORM
The pulse signal is generated while the gate signal is high.
Manual operation: See "Gate Input Polarity" on page 117
[:SOURce<hw>]:PULM:TRIGger:EXTernal:IMPedance <Impedance>
Selects the impedance for external pulse trigger.
Parameters:
<Impedance> G50 | G10K
*RST: G50
Example:
SOUR:PULM:TRIG:EXT:IMP G50
selects 50 Ohm as the trigger impedance for the external pulse
trigger.
Manual operation: See "External Impedance" on page 114
[:SOURce<hw>]:PULM:TRIGger:EXTernal:SLOPe <Slope>
Sets the polarity of the active slope of an externally applied trigger signal.
Parameters:
<Slope> NEGative | POSitive
*RST: POSitive
Example:
PULM:TRIG:EXT:SLOP NEG
The pulse generator is triggered on the negative slope of the
external trigger signal.
Manual operation: See "Ext. Trigger Input Slope" on page 117
[:SOURce<hw>]:PULM:TRIGger:MODE <Mode>
Selects the trigger mode for pulse modulation.
Parameters:
<Mode> AUTO | EXTernal | EGATe
AUTO
The pulse modulation is generated continuously.
EXTernal
The pulse modulation is triggered by an external trigger event.
The trigger signal is supplied via the trigger connector.
SOURce:PULM Subsystem