Total Station Manual
This will help screw A to adjust the inclination of the instrument.
In Tools Menu, choose “1 Adjust”, and then press [F3] to enter “Adjust Tilt X”.
These are the calibration of x-direction of compensator’s vertical axis.
○
1 Level instrument, focus
on the reticle of collimator,
record the vertical angle V0.
Use fine tuning to set
vertical angle to V0+3’,focus
on the reticle center
accurately, wait for stable
value, press [F4](OK).
○
2 Use fine tuning to set the
vertical angle to V0-3’, focus
on the reticle center
accurately, wait for stable
value, press [F4] (OK).
○
3
Use fine tuning to set the
vertical angle as V0,focus on
the reticle center accurately.
○
4
Reverse the telescope,
use face right to focus on the
reticle of collimator, record
the vertical angle V1.Use fine
tuning to set the vertical
angle as V1-3’,focus on the
reticle center accurately, wait
for stable value, pressF4(OK).