Total Station Manual
○
2 Use fine tuning to set the
vertical angle to V0-3’, focus
on the reticle center
accurately, then turn the
instrument counterclockwise
90 °, wait for stable value,
press [F4] (OK) ,and then turn
90 ° clockwise back to the
original direction.
○
3
Use fine tuning to set the
vertical angle as V0,focus on
the reticle center accurately.
○
4
Reverse the telescope,
use face right to focus on the
reticle of collimator, record
the vertical angle V1.Use fine
tuning to set the vertical
angle as V1-3’,focus on the
reticle center accurately, then
turn the instrument
counterclockwise 90 °, wait
for stable value,
pressF4(OK) ,and then turn
90 ° clockwise back to the
original direction.
○
5
Use fine tuning to set the
vertical angle as V1+3’, focus
on the reticle center
accurately, then turn the
instrument counterclockwise
90 °, wait for stable value,
press [F4](OK).