5.5 Shutdown and Startup Settings
5-17
CBM-20A/20Alite
5
4
Input the column temperature.
This item can be set to 0.0 or to a temperature in the range 4.0 to 80.0 (ï‚°C). If 0.0
is set, the column turns OFF.
5
To execute pumping during cooldown at the concentration and at half the flow rate
specified for the pump in the method file, click the [Set flow rate at half of that in method
during cooldown.] check box.
6
Click the check boxes for required items under [Step 2: System Shutdown].
Pumping operation by the solvent delivery module, temperature control of the
oven, and the detector lamps are all turned OFF when shutdown is executed.
7
Click .
The settings are saved and the window closes.
The cooldown-time setting and the settings made under "Step 2: System Shutdown"
are all reflected in the "Startup/Shutdown Setting" section.