3. Preparation
3-12
CBM-20A/20Alite
Allocation with [Auto] Selected
The pumps are given letters A, B, C, and D in order of increasing link address. The detector
and oven names are also assigned in the same way. If the CBM-20Alite is incorporated in an
LC-30A/20A-series solvent delivery module, this delivery unit will always be pump A.
Allocation with [Fixed] Selected
Select [Fixed] to fix the relationship between all the components in the system. For example,
with [Auto] selected, if there are 2 detectors and the power to detector A is turned OFF, then
detector B is automatically recognized as detector A. In the same way, if there are 3 pumps
used and the power to pump A is turned OFF, then pumps B and C are automatically
recognized as pumps A and B. This makes it necessary to change existing parameter and
program settings.
With [Fixed] selected, however, detector B continues to be recognized as detector B even if
the power to detector A is turned OFF. In the same way, pumps B and C continue to be
recognized as pumps B and C even if the power to pump A is turned OFF. The oven names
are also assigned in the same way as pumps or detectors.
The default setting is [Auto].
Allocation when power to Pump.A and DET.A is turned OFF during control
Fig. 3.2
During control of all components
Auto Fixed
PUMP.A
ON
CBM DET.A
ON
PUMP.B
ON
PUMP.C
ON
DET.B
ON
Auto
OFF
CBM
OFF
PUMP.A
ON
PUMP.B
ON
DET.A
ON
Fixed
OFF
CBM
OFF
PUMP.B
ON
PUMP.C
ON
DET.B
ON