EasyManua.ls Logo

Shimadzu CBM-20A - Creating Sequence Tables

Shimadzu CBM-20A
412 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. Basic Operation
4-26
CBM-20A/20Alite
4.3.3 Creating Sequence Tables
Create a sequence table using a sequence file. Input analysis conditions, such as vial
numbers and injection volumes, in the sequence table.
1
If the connected autosampler is the SIL-30A/20A series, input the rack number in the
[Rack No.] column and press the [Tab] key.
The default settings [1], [1], [1], [10], [-], and [1.00] are displayed respectively in the
[From] and [To] columns under [Sample No.], the [Injections/Vial] column, the [Injection
Volume] column, the [Method] column, and the [Run Time] column.
The default setting for [Injection Volume] varies depending on the connected
autosampler.
If the connected autosampler is the SIL-10ADvp/10A/10Ai/10AXL/10AF/10AP, no
value is displayed in the [Rack No.] column.
It is possible to set [-] in the [Rack No.] and [From] columns. In this case, analysis
operation is performed in accordance with the [Method] setting for that line but
injection operation is not performed by the autosampler.
2
Make corrections as required.
When using a time program for analysis, select the required method file from the pull-
down menu. When a method file is selected, the setting shown in the [Run Time]
column changes to [-]. When a method file is not selected, enter the analysis time in the
[Run Time] column.
Two or more method files can be selected for one analysis sequence.
3
The input method for line 2 and subsequent lines is the same as for line 1.
Note that the values displayed in the [To] column under [Sample No.], the [Injection/
Vial] column, the [Injection Volume] column, the [Method] column, and the [Run Time]
column are the same as the input values for the previous line.
Up to 100 lines can be input in a sequence table. The sequence is composed in
order starting from line 1. It is not possible to skip lines.
4
When the required settings have been made in the sequence table, the table is saved
as a sequence file.

Table of Contents

Other manuals for Shimadzu CBM-20A

Related product manuals