9. Technical Information
9-2
CBM-20A/20Alite
9.1 Installation
9.1.1 Installation Site
Suitable Sites and Preparation
To ensure safe operation, install the instrument in a suitable location that satisfies the following conditions.
WARNING
• Ample ventilation
The solvents used with the HPLC system are often flammable and toxic.
Therefore, the room where the instrument is installed must be well-ventilated.
• No fire sources used near the instrument
The solvents used with the HPLC are often flammable. Therefore, the use of open flame where the
instrument is installed must be strictly prohibited. Also, do not install in the same room with equipment
that emits or could potentially emit sparks.
• Fire extinguishers permanently available
Have fire extinguishers permanently available in case of fire.
• Protective equipment provided near the instrument
If solvent gets into the eyes or onto the skin, it must be flushed away immediately.
Provide equipment, such as eye wash stations and safety showers, as close to the instrument as
possible.
CAUTION
• Avoid dust or corrosive gas
To ensure a long service life of the instrument and preserve its performance levels, avoid installing it in
places subject to large amounts of dust or corrosive gas.
• Keep away from equipment generating strong magnetic fields
To ensure proper operation, do not install the instrument in places subject to strong magnetic fields.
If the power supply line is subject to high electrical noise, install a surge protector.
• Install the instrument in the location that satisfies the following conditions to preserve the
performance:
• room temperature is between 4 and 35 ºC, with minimal temperature variation through a day.
• air currents from heating or air conditioning equipment are not directed on the instrument.
• sunlight does not shine directly on the instrument.
• there is no vibration.
• humidity stays within 20 - 85 %.
• place without condensation