9.1 Installation
9-3
CBM-20A/20Alite
9
Required Installation Space
Typical system configurations and required installation spaces are shown in the figures below.
Fig. 9.1
CAUTION
• The weight of CBM-20A is 5.5 kg. During installation, consider the entire weight combined
with other LC components.
The lab table on which this instrument is installed should be strong enough to support the total weight
of the LC system. It should be level, stable and have depth of at least 600 mm.
If these precautions are not followed, the instrument could tip over or fall off the table.
• Keep at least 100 mm between the rear of the instrument and the wall.
This allows for sufficient air circulation to provide cooling and prevent the instrument from overheating
and impairing the performance.
System 1 (with Manual Injector)
System 2 (with Autosampler)
(Unit : mm)