7.4 Equipment Required for Validation
7-7
CBM-20A/20Alite
7
Standard Reagents for Validation
A list of standard reagents required for validation is shown below. The customer should
prepare standard reagents to the stated specifications.
Standard sample Part No. Description
Caffeine set
(5 concentrations)
S228-45725-91
For inspection of the absorbance linearity for the UV-VIS
spectrophotometric and photodiode array detectors.
For also inspection of system reproducibility for a system
equipped with a UV-VIS spectrophotometric or photodiode
array detector.
Caffeine (250 mg/L) S228-45725-06
For inspection of system reproducibility for a system
equipped with a refractive index detector, inspection of
autosampler carry-over, and inspection of the gradient
concentration accuracy for gradient systems.
Naphthalene (60 mg/
L)
S228-32996-01
For inspection of system reproducibility for a system
equipped with a spectrofluorometric detector.
Glycerol (0.872 mg/L) S228-32996-05 For inspection of the span for the refractive index detector.