7. Hardware Validation
7-8
CBM-20A/20Alite
Hardware Testing Supplies
A list of supplies required for hardware validation is shown below. Note that items such as
autosampler vials or mobile phase solutions may be required in addition to the items listed.
Implement Part No. Description
Resistor tube S228-45726-91
I.D. 0.13 mm × 2 m + I.D. 0.8 mm × 2 m
For inspection of flow rate and gradient concentration
accuracy for solvent delivery module, etc.
Syringe
S046-00001
or
S046-00038-01
For inspection of the absorbance linearity for the UV-VIS
spectrophotometric and photodiode array detectors.
For also inspection of the span for the refractive index
detector.
This item is provided with detectors as a standard
accessory.
Syringe adapter S228-15672-91 Same as above.
Coupling 1.6C S228-16004-13
For each kind of inspection and in plumbing the detector.
This item is provided with each component as a standard
accessory.
Male nut, PEEK S228-18565 Same as above.
Plug S228-16006
For inspection of the drift/noise for the refractive index
detector.
Low-pressure
Hg (Mercury) lamp set
S200-38423
For inspection of the wavelength accuracy for the UV-VIS
photodiode array detector and the spectrofluorometric
detector.
Hg (Mercury) lamp
holder
S228-34170-91
For inspection of the wavelength accuracy for the UV-VIS
photodiode array detector.
S228-34478-91
For inspection of the wavelength accuracy for the
spectrofluorometric detector.
PTFE block assembly S228-34319-91
For inspection of the wavelength accuracy for the
spectrofluorometric detector.
Column
Shim-pack VP-ODS
or
LUNA C18(2)
S228-34937-91
or
00F-4252-E0
Particle size: 5µm
Column Dimension: I.D. 4.6 mm × length 150 mm
(An equivalent ODS column may also be used.)
For the system validation.