5. Application Operation
5-130
CBM-20A/20Alite
5.13.4 Detector
SPD-20A
SPD-20AV
Module Function Description
Setting
range
Minimum
unit
Default
setting
Remark
DetectorA
DetectorB
WAVE Wavelength 1 (nm)
190 to 700 1 254
Setting for single
mode
190 to 370 1 254
Settings for dual
mode
371 to 700 1 500
WAVE2 Wavelength 2 (nm)
190 to 370 1 254
371 to 700 1 500
RESP Response 0 to 10 1 4
LAMP Lamp selection 0:OFF, 1:D2 1 1
CELL.T Cell temperature (ºC) 0, 9 to 50 1 40 0 to OFF
ZERO
Execution of auto-
zero
SV.WASTE
Solvent recycle valve
switching
0: According
to the SV
level setting
1: Always
waste side
10
Setting for solvent
recycle valve
Module Function Description
Setting
range
Minimum
unit
Default
setting
Remark
DetectorA
DetectorB
WAVE Wavelength 1 (nm)
190 to 370 1 254
Setting for D2
lamp
371 to 700 1 500 Settings for W
lamp in single
mode
701 to 900 1 500
190 to 370 1 254
Settings for D2
and W lamps in
single mode
371 to 700 1 254
701 to 900 1 254
371 to 700 1 500
Settings for W
lamp in dual mode
701 to 900 1 800
190 to 370 1 254
Settings for D2
and W lamps in
dual mode
371 to 700 1 500
701 to 900 1 800