7. Hardware Validation
7-10
CBM-20A/20Alite
5
Click the [Configuration] tab and check the version number of the system controller
(version number of the control unit) displayed in the configuration table.
7.5.3 System Control
Objective
Check control operation of the system controller.
Check Procedure
1
Start up the system controller.
2
Start up Internet Explorer on a PC and specify the URL of the system controller.
3
Enter the analysis sequence as shown in the following table.
4
Enter the analysis method as shown in the following table.
• CHECK
CRITERIA
• The version numbers for both the contents and the
control unit are the same as the administrated ones.
Rack No.
Sample
No. From
Sample
No. To
Injections/
Vial
Injection
Volume
Method Run Time
1 1 1 1 10 - 1.00
System Settings
Isocratic system
Click in the "Analysis" tab page.
Set the flow rate for ISO1 to 0.5 mL/min.
Low-pressure gradient system
Click in the "Analysis" tab page.
Set the flow rate for LP.GE1 to 1 mL/min and
concentration B to 50 %.
High-pressure gradient system
Click in the "Analysis" tab page.
Set the flow rate for B.GE1 to 1 mL/min and
concentration B to 50 %.