7.6 System Validation
7-15
CBM-20A/20Alite
7
Check Procedure
1
Check that the pumping mode in the "Configuration" tab page is set to [ISO/Binary
Pump]. If it is not, select [ISO/Binary Pump] from the pull-down menu.
For details on the display of the "Configuration" tab page, refer to "3.4 Determine the
System Configuration" P. 3-7.
2
Enter the parameters for the solvent delivery module, autosampler, column oven, and
detector in the method file as shown in the following table.
For details on setting component parameters, refer to "4.2 Creating Method Files" P. 4-10.
Parameters Set value
Solvent delivery
module (Pump)
Flow Rate 1 mL/min
Maximum Pressure 20.0 MPa
Autosampler
Sampling Speed 15 µL/sec
Rinse Volume 200 µL
Rinse Speed 35 µL/sec
Rinse Mode No Rinse
Column oven Oven Temperature 40 ºC
Detector
Wavelength 272 nm
AUX Range 1 AU/V
Response 0.5 sec