1 IMPORTANT INFORMATION
Recognizing malfunctions
Every deviation from normal operation is to be regarded as a serious indication of a func-
tional impairment. These are, amongst others:
● Warning displays
● Significant drifts in measured results
● Increased power consumption
● Higher temperatures of system components
● Monitoring devices triggering
● Smells or smoke emission
● Heavy contamination.
Avoiding damage
Electrical connection
Ensure the device can be switched off with a power isolating switch/circuit breaker in accor-
dance with EN 61010-1.
NOTE:
In order to avoid malfunctions that can cause direct or indirect personal injury
or property damage, the operator must ensure:
▸ The responsible maintenance personnel are present at any time and as fast
as possible
▸ The maintenance personnel are adequately qualified to react correctly to
malfunctions of the measuring system and any resulting operational inter-
ruptions (e.g., when used for measurement and control purposes)
▸ The malfunctioning equipment is switched off immediately in case of doubt
and that switching off does not cause collateral malfunctions.