EasyManuals Logo

SICK DUSTHUNTER T100 User Manual

SICK DUSTHUNTER T100
261 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #78 background imageLoading...
Page #78 background image
78
8012428/YWL2/3-0/2016-08| SICKO P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER T
Subject to change without notice
4 START-UP AND PARAMETER SETTINGS
4.4.5 Setting the analog inputs parameters
Select the “Configuration / I/O Configuration / Input Parameters DUSTHUNTER” directory
to set the analog inputs.
Fig. 52: SOPAS ET menu: MCU/Configuration/IO configuration/Input Parameters
Field Parameter Remark
Temperature
Source
Constant Value A fixed value is used to calculate the scaled value.
This parameter opens the “Constant Temperature” field to enter the scaling value in °C (° F for
imperial units) or K.
Analog Input 1 The value from an external sensor connected to analog input 1 (standard scope of delivery) is
used to calculate the scaled value.
This parameter opens the “Analog input 1 - Temperature” field to set the lower and upper range
limit values and the Live Zero value.
Pressure
Source
Constant Value A fixed value is used to calculate the scaled value.
This parameter opens the “Constant Pressure” field to enter the scaling value in mbar (= hPa).
Analog Input 2 The value from an external sensor connected to analog input 2 (standard scope of delivery) is
used to calculate the scaled value.
This parameter opens the “Analog input 2 - Pressure” field to set the lower and upper range limit
values and the Live Zero value.
Moisture
Source
Constant Value A fixed value is used to calculate the scaled value.
This parameter opens the “Constant Moisture” field to enter the scaling value in %.
Analog Input 3 The value from an external sensor connected to analog input 3 (optional module required) is
used to calculate the scaled value.
This parameter opens the “Analog input 3 - Moisture” field to set the lower and upper range
limit values and the Live Zero value.
Oxygen
Source
Constant Value A fixed value is used to calculate the scaled value.
This parameter opens the “Constant Oxygen” field to enter the scaling value in %.
Analog Input 4 The value from an external sensor connected to analog input 4 (optional module required) is
used to calculate the scaled value.
This parameter opens the “Analog input 4 - Oxygen” field to set the lower and upper range limit
values and the Live Zero value.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK DUSTHUNTER T100 and is the answer not in the manual?

SICK DUSTHUNTER T100 Specifications

General IconGeneral
BrandSICK
ModelDUSTHUNTER T100
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals