5 Calibration
25
Q Start time
Note
In selecting the AIC based on the internal clock, it is required to set the start time.
The following is the procedure to set the next start time.
1.
Pressing the the right button of [START] in the AIC SET screen will display the AIC
START screen.
Fig. 19 AIC START screen
2. Pressing the button showing the value to be changed will reverse the display
mode. Change the value by the [T] and [S] button, and press the [SET] key.
3. The instrument changes the setting and the screen returns to the AIC SET screen.
Note
z The start time setting is based on the internal clock.
z It is not available to change into an impossible date.
When this happens, the existing date closest to that before the set date is automatically selected.
z When the start time previous to the current time is set, the setting value of start time is changed
into the minimum value which comes in future by adding multiple number of the interval set
values.
z When the start time set value becomes a previous time by changing the internal clock setting
(refer to " 6.5.2 Time adjust " (page 48)), the start time is changed into the minimum value which
comes in future by adding multiple number to the current time set value.
z When the AIC start time is set later than 2090/1/1/0:00, the AIC does not start.
When the “Battery” alarm occurs, confirm the setting time.
Setting item Setting range
Year 2000 to 2090
Month 1 to 12
Day 1 to 31
Hour 0 to 23
Minute 0 to 59
Year
Month
Day
Hour
Minute