8 Troubleshooting
79
8.2 Troubleshooting
This chapter describes the troubleshooting mainly for the parts replacement advices and
check items for customers. 
If these counteractions do not solve the problems, be sure to contact HORIBA.
Note
Before trying the counteractions, check the followings again:
z Power is connected
z Power voltage/capacity meet the specification
z Parts replacement is correctly performed
Q No indication
Q No/ low output signals
Cause Check item Counteraction Reference
Back light is extinguished
-
Touch the screen.
page 16
page 47
Analyzer power is OFF Confirm the power switch. Turn the analyzer power ON. page 3
Others
-
Contact HORIBA.
-
Cause Check item Counteraction Reference
Analog output mode has 
been selected
Confirm if the screen is set to analog 
output mode.
Exit from analog output screen to 
Meas. mode.
page 38
The output is hold. Confirm the delivery specification.
--
Analyzer power is OFF Confirm the power switch. Turn ON the analyzer power. page 3
Tube is disconnected.
Confirm that the sample tubes are not 
disconnected or broken.
Adjust and fix the defective tubes.
-
Pump is stopped.
Confirm the flow by the flowmeter on 
the analyzer front panel.
Confirm the pump power switch.
Turn ON the pump power. page 3
Leakage from primary filter
Confirm that the filter element is 
correctly attached.
Confirm if the O-ring or packing has no 
deterioration.
Ensure the correct attachment.
Replace the O-ring and packing.
page 61
Leakage from mist catcher 
attachment part
Confirm that the mist catcher is 
correctly attached.
Confirm if there is no crack in the 
rubber joint.
Attach it again.
Replace the rubber joint.
page 64
Leakage from the secondary 
filter attachment part
Confirm that the secondary filter is 
correctly attached.
Confirm if there is no missing O-ring or 
packing to be attached.
Attach it again. page 62
Leakage from NOx converter 
attachment part 
(With NOx converter)
Confirm that the NOx converter is 
correctly attached.
Confirm there is no crack in the rubber 
joint.
Attach it again.
Replace the rubber joint.
page 67
Sample outlet is clogged Check the sample outlet part.
Either release it into air or prepare 
tubing layout free from 
backpressure.
-
Specified flow: (0.6 L/min ± 
0.1 L/min) is not flowed
Confirm the flow by the flowmeter on 
the analyzer front panel.
Adjust the flow to the specified rate 
by the flow adjusting needle valve.
page 30