8012428/YWL2/3-0/2016-08| SICKO P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER T
Subject to change without notice
4 START-UP AND PARAMETER SETTINGS
4.1.4 Connection to the device via Ethernet (option)
Recommended procedure:
1 The MCU must be switched off.
2 Connect the MCU with the network.
3 Connect the PC to the same network.
4 Switch the MCU on.
5 Start SOPAS ET.
6 Search settings”
7 “Device family oriented search”
8 Click the desired MCU
9 Make the settings:
Ethernet communication (always clicked)
USB communication (always clicked)
Serial communication: Do not click
10 Specify the IP addresses
IP address: see “Setting the Ethernet module parameters”, page 85
11 Do not click a COM port
12 Assign a name for this search
13 “Finish”
The Ethernet interface module (see “Options for MCU control unit”, page 125) must be
installed in the MCU (see “Fitting the interface and I/O module (option)”, page 56) and
configured (see “Setting the Ethernet module parameters”, page 85) to connect to the
measuring system via Ethernet.