EasyManua.ls Logo

SICK DUSTHUNTER T100

SICK DUSTHUNTER T100
261 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Optional Functions
97
Q Interval
Sets the time span (interval) for the purge start.
1. Pressing the button at the right of [INTERVAL] in the PURGE SET screen displays
the PURGE INTERVAL screen.
Fig. 71 PURGE INTERVAL screen
2. Pressing the button which setting is to be changed will reverse the value display.
Change the value by the [T] and [S] button and press the [SET] key.
3. The setting is changed, and the screen returns to the PURGE SET screen.
Note
The recommended purge interval is 12 hours. When the above recommended time does not meet the
actual condition, change the setting after confirmation.
Q In blow time
Set the IN BLOW TIME in purge.
1. Pressing the button at the right of [IN BLOW TIME] in the PURGE SET screen
displays the PURGE IN BLOW TIME screen.
Fig. 72 PURGE IN BLOW TIME screen
Setting item Setting range
Hour 1 to 99

Table of Contents

Related product manuals