6 TROUBLESHOOTING
6.3.3 Replacing the fuse
▸ Control unit MCU must be disconnected from the power supply and potential-free.
▸ Open the door of the MCU, remove and open fuse holder (1).
▸ Replace the defective fuse (2) (see “Miscellaneous”, page 125).
▸ Close and attach the fuse holder.
▸ Close the door and connect power voltage.